See the components you need for the Attest™ eBowie-Dick Test System.
Attest ebowie-dick auto-reade 1190
Attest™ eBowie-Dick Test Card 10135, 10135B
Attest™ eBowie-Dick Card Holder 10135CH, 10135CHB
Attest™ eBowie-Dick module within Attest™ Connect PC Application
Reusable holder for Attest ebowie-dick test card 10135.
Can be recycled if programs exist for this product in your area.
Suggested Applications
The Attest ebowie-dick test card 10135 and Attest eBowie-Dick Card Holder 10135CH are used in conjunction with the Attest™ eBowieDick Auto-reader 1190 to monitor the equipment performance of steam sterilisers each day, as well as after a steam steriliser repair
The Attest ebowie-dick test card 10135 and Attest ebowie-dick card holder 10135CH is designed for testing air removal efficiency of 132°C (270°F), 134°C (273°F), and 135°C (275°F) for 3.5 minute test cycles in dynamic air removal steam sterilisers with chamber volumes greater than 54L (approximately 2 cubic feet).